Z-2075-2012 Class II Terminated

Recalled by Hospira Inc. — Lake Forest, IL

FDA device recall Z-2075-2012 was initiated by Hospira Inc. on Jul 13, 2012 and is designated Class II. Reason for recall: The float valve in the burette sticks to the burette wall and does not open or close properly. The recall status is terminated (terminated Nov 28, 2016). Affected quantity: 4,420 sets.

Recall Details

Product Type
Devices
Report Date
Aug 1, 2012
Initiation Date
Jul 13, 2012
Termination Date
Nov 28, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,420 sets

Product Description

Hospira LifeShield Latex-Free 100 mL Burette Set, Convertible Pin, 77 inch with 2 Prepierced Injection Sites and Option-Lok Microdrip Soluset with Calibrated Burette and Precision Drip Chamber; an Rx sterile medical device used to administer fluids; 20 individually wrapped sets per case; Made in Costa Rica, Hospira, Inc., Lake Forest, IL 60045 USA; list 12722-65. Intended use: for the administration of fluids.

Reason for Recall

The float valve in the burette sticks to the burette wall and does not open or close properly.

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of California, Colorado, Connecticut, Florida, Illinois, Kentucky, Maryland, Massachusetts, Minnesota, Oklahoma, Texas, Utah, Virginia, Washington, West Virginia and Wyoming, and the countries of Barbados and Canada.

Code Information

list 12722-65, lot number 96-098-5H

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2075-2012
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.