Z-2077-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 13, 2018
- Initiation Date
- March 5, 2018
- Termination Date
- August 31, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5293
Product Description
AIA-PACK TSH 3rd Generation; Part Number: 020294 Assay, Thyroid
Reason for Recall
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.
Code Information
All Lot Codes