Z-2086-2018 Class II Terminated

Recalled by Tosoh Bioscience Inc — Grove City, OH

FDA device recall Z-2086-2018 was initiated by Tosoh Bioscience Inc on Mar 5, 2018 and is designated Class II. Reason for recall: Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay. The recall status is terminated (terminated Aug 31, 2020). Affected quantity: 9033.

Recall Details

Product Type
Devices
Report Date
Jun 13, 2018
Initiation Date
Mar 5, 2018
Termination Date
Aug 31, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9033

Product Description

ST-AIA PACK FSH; Part Number: 025265 Assay, Reproductive Hormone

Reason for Recall

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

Distribution Pattern

Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.

Code Information

All Lot Codes

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2086-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.