Z-2086-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 13, 2018
- Initiation Date
- March 5, 2018
- Termination Date
- August 31, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9033
Product Description
ST-AIA PACK FSH; Part Number: 025265 Assay, Reproductive Hormone
Reason for Recall
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.
Code Information
All Lot Codes