Z-2092-2012 Class II Terminated

Recalled by Zimmer Inc. — Warsaw, IN

FDA device recall Z-2092-2012 was initiated by Zimmer Inc. on Apr 30, 2012 and is designated Class II. Reason for recall: The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design… The recall status is terminated (terminated Jun 4, 2014). Affected quantity: 65 units.

Recall Details

Product Type
Devices
Report Date
Aug 8, 2012
Initiation Date
Apr 30, 2012
Termination Date
Jun 4, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
65 units

Product Description

Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.

Reason for Recall

The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall

Distribution Pattern

Worldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.

Code Information

Part 00-5901-026-40 Lot 62011774, 61870067, and 61813699 Initially the 00-5901-026-40 replacement jaws will remain available for order until all of the alternative instruments are in place, however they will also be required to be returned. The final lot numbers and manufacture dates will be provided in a subsequent status update.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2092-2012
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.