Z-2117-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 13, 2018
- Initiation Date
- April 6, 2018
- Termination Date
- June 30, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 100
Product Description
Rival View Kit Locking and Non-Locking Screw D 2.7MM, D 3.2MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-009C
Reason for Recall
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Distribution Pattern
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Code Information
UDI: 18056099648578, Lot: B1094193, B1104809, B1107372