Z-2124-2016 Class II Terminated

Recalled by Esaote S.p.A. — Firenze, N/A

Recall Details

Product Type
Devices
Report Date
July 13, 2016
Initiation Date
June 6, 2016
Termination Date
May 30, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11 USA; 29 OUS

Product Description

Magnetic Resonance System G-scan Brio (Part Number /Model No.101002000)

Reason for Recall

The device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table. Additionally, a weight limit is added to the patient seat for weight-bearing examinations.

Distribution Pattern

Domestic: FL, TX, OH, MN, NV; Foreign: No customers in Canada; No VA/DOD.

Code Information

Item No., Unit Description and Serial No.s: 101002000, UNIT, G-SCAN BRIO, 05202; 101002000, UNIT, G-SCAN BRIO, 05203; 101002000, UNIT, G-SCAN BRIO 2.1A; 05205; 101002000, UNIT, G-SCAN BRIO 2.1A, 05207; 101002000 UNIT, G-SCAN BRIO 2.1A, 05208; 101002000 UNIT, G-SCAN BRIO 2.1A, 05209; 101002000 UNIT, G-SCAN BRIO 2.1A, 05210; 101002000 UNIT, G-SCAN BRIO 2.1A, 05211; 101002000 UNIT, G-SCAN BRIO 2.1A, 05305; 101002000 UNIT, G-SCAN BRIO 2.1A, 05308