Z-3053-2020 Class II Terminated

Recalled by ESAOTE S.P.A — Genova, N/A

FDA device recall Z-3053-2020 was initiated by ESAOTE S.P.A on Sep 8, 2020 and is designated Class II. Reason for recall: The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the… The recall status is terminated (terminated Apr 17, 2023). Affected quantity: 989 total adapters in the US.

Recall Details

Product Type
Devices
Report Date
Oct 7, 2020
Initiation Date
Sep 8, 2020
Termination Date
Apr 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
989 total adapters in the US

Product Description

AC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)

Reason for Recall

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

Distribution Pattern

US Nationwide

Code Information

Serial Numbers: Below K17350001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3053-2020
openFDA file date
Nov 3, 2026
Loaded by RegDataLab
Nov 4, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-3053-2020. Report a data issue: hello@regdatalab.com.