Z-0206-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 31, 2018
- Initiation Date
- June 11, 2018
- Termination Date
- May 12, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10
Product Description
MyLab Seven, Model 6400
Reason for Recall
The probe power monitoring settings have been found to be not appropriate under specific operating conditions when connected to the devices listed below. This issue may result in overheating of the probe head. The probe power monitoring settings may not be effective in specific operating conditions so that the surface of the probe head may overheat in around one minute of continuous operations. The patient may experience discomfort, reddening of epidermis and/or mild pain. To date, no patient injury has been reported with regards to this issue.
Distribution Pattern
Worldwide distribution - US Nationwide in the states of AK, AZ, CA, IN, MO, NM, PA, SC, and TX. and countries Argentina and Australia Austria Bangladesh Belgium Brazil Bulgaria China Costa Rica Denmark Egypt Finland France Germany Greece Hong Kong Hungary Iceland India Iran Italy Japan Kuwait Mexico Netherlands Norway Paraguay Poland Portugal Russian Fed. Saudi Arabia Singapore Slovenia South Africa South Korea Spain Switzerland Taiwan Thailand Tunisia Turkey United Kingdom Utd.Arab Emir.
Code Information
n/a