Z-0206-2019 Class II Terminated

Recalled by ESAOTE S.P.A. — Sesto Fiorentino, N/A

Recall Details

Product Type
Devices
Report Date
October 31, 2018
Initiation Date
June 11, 2018
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product Description

MyLab Seven, Model 6400

Reason for Recall

The probe power monitoring settings have been found to be not appropriate under specific operating conditions when connected to the devices listed below. This issue may result in overheating of the probe head. The probe power monitoring settings may not be effective in specific operating conditions so that the surface of the probe head may overheat in around one minute of continuous operations. The patient may experience discomfort, reddening of epidermis and/or mild pain. To date, no patient injury has been reported with regards to this issue.

Distribution Pattern

Worldwide distribution - US Nationwide in the states of AK, AZ, CA, IN, MO, NM, PA, SC, and TX. and countries Argentina and Australia Austria Bangladesh Belgium Brazil Bulgaria China Costa Rica Denmark Egypt Finland France Germany Greece Hong Kong Hungary Iceland India Iran Italy Japan Kuwait Mexico Netherlands Norway Paraguay Poland Portugal Russian Fed. Saudi Arabia Singapore Slovenia South Africa South Korea Spain Switzerland Taiwan Thailand Tunisia Turkey United Kingdom Utd.Arab Emir.

Code Information

n/a