Z-2131-2018 Class II Terminated

Recalled by Orthofix, Inc — Lewisville, TX

FDA device recall Z-2131-2018 was initiated by Orthofix, Inc on Apr 6, 2018 and is designated Class II. Reason for recall: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not… The recall status is terminated (terminated Jun 30, 2021).

Recall Details

Product Type
Devices
Report Date
Jun 13, 2018
Initiation Date
Apr 6, 2018
Termination Date
Jun 30, 2021
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

MTP Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-B-01001); b) Medium (002-B-02002); c) Large (002-B-03003)

Reason for Recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Distribution Pattern

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Code Information

(UDI)/Lot: a) (18056099647557) B1098799, B1115575; (18056099647564) B1113628; (18056099647571) B1098800