Z-2141-2024 Class II Ongoing

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

FDA device recall Z-2141-2024 was initiated by Fresenius Medical Care Holdings, Inc. on May 16, 2024 and is designated Class II. Reason for recall: Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded")… The recall status is ongoing. Affected quantity: 4 units.

Recall Details

Product Type
Devices
Report Date
Jun 19, 2024
Initiation Date
May 16, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Reason for Recall

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

Distribution Pattern

US Nationwide distribution in the states of AL, IN, AK, WI.

Code Information

UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087