Z-2171-2018 Class II Terminated

Recalled by BioMerieux SA — Marcy L'Etoile, N/A

Recall Details

Product Type
Devices
Report Date
June 20, 2018
Initiation Date
March 23, 2018
Termination Date
June 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 boxes COLOR GRAM2 with 4 - 240ML bottles in each

Product Description

Color Gram 2 (COLOR GRAM 2 - F), these stains are used to stain bacterial and fungal microorganisms for direct examination, according to the Gram method. Dye and chemical solution stains for medical purposes are mixtures of synthetic or natural dyes or nondye chemicals in solutions used in staining cells and tissues for diagnostic histopathology, cytopathology, or hematology.

Reason for Recall

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Distribution Pattern

International distribution.

Code Information

Color Gram 2 (COLOR GRAM 2 - F) (ref: 55542) Lot number G64508