Z-2207-2020 Class II Terminated

Recalled by Nobel Biocare Usa Llc — Yorba Linda, CA

FDA device recall Z-2207-2020 was initiated by Nobel Biocare Usa Llc on Jul 19, 2019 and is designated Class II. Reason for recall: The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm)… The recall status is terminated (terminated Oct 27, 2020). Affected quantity: 221 implants.

Recall Details

Product Type
Devices
Report Date
Jun 10, 2020
Initiation Date
Jul 19, 2019
Termination Date
Oct 27, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
221 implants

Product Description

NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Reason for Recall

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

Distribution Pattern

US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None

Code Information

Lot number: 13080792

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2207-2020
openFDA file date
Nov 3, 2026
Loaded by RegDataLab
Nov 4, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2207-2020. Report a data issue: hello@regdatalab.com.