FDA device recall Z-1851-2020 was initiated by Nobel Biocare Usa Llc on August 28, 2019 and is designated Class II. Reason for recall: The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the … The recall status is terminated (terminated April 28, 2021). Affected quantity: 228 pieces.
Recall Details
- Product Type
- Devices
- Report Date
- May 13, 2020
- Initiation Date
- August 28, 2019
- Termination Date
- April 28, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 228 pieces
Product Description
NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.
Reason for Recall
The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).
Distribution Pattern
International distribution - the country Russia only.
Code Information
Lot:12129104