Z-0658-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 2, 2022
- Initiation Date
- December 17, 2021
- Termination Date
- October 25, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 702 pieces
Product Description
NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R
Reason for Recall
Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.
Distribution Pattern
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.
Code Information
Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497