Z-0658-2022 Class II Terminated

Recalled by Nobel Biocare Usa Llc — Yorba Linda, CA

Recall Details

Product Type
Devices
Report Date
March 2, 2022
Initiation Date
December 17, 2021
Termination Date
October 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
702 pieces

Product Description

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reason for Recall

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

Code Information

Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497