Z-2269-2012 Class III Terminated

Recalled by Baxter Healthcare Corp. — Round Lake, IL

FDA device recall Z-2269-2012 was initiated by Baxter Healthcare Corp. on Mar 4, 2011 and is designated Class III. Reason for recall: Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The… The recall status is terminated (terminated Aug 27, 2012). Affected quantity: 19,380 units.

Recall Details

Product Type
Devices
Report Date
Sep 5, 2012
Initiation Date
Mar 4, 2011
Termination Date
Aug 27, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19,380 units

Product Description

Mini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an active time scale and rate selector switch; product codes 2M8171 and 2M8171R. Product Usage: These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician. Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication.

Reason for Recall

Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum

Distribution Pattern

45,865 units

Code Information

product codes 2M8171 and 2M8171R, all serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2269-2012
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.