Z-2276-2018 Class II Terminated

Recalled by Philips Medical Systems (Cleveland) Inc — Cleveland, OH

FDA device recall Z-2276-2018 was initiated by Philips Medical Systems (Cleveland) Inc on Mar 28, 2018 and is designated Class II. Reason for recall: A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or… The recall status is terminated (terminated Apr 21, 2020). Affected quantity: 28 total.

Recall Details

Product Type
Devices
Report Date
Jul 4, 2018
Initiation Date
Mar 28, 2018
Termination Date
Apr 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28 total

Product Description

Ingenuity TF PET/CT 882442

Reason for Recall

A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.

Distribution Pattern

The systems were distributed to the following US states: HI, KS, and MO. The systems were distributed to the following foreign countries: Algeria, Austria, Brazil, China, Indonesia, Japan, Kenya, Korea, Mexico, Poland, Turkey, and Uganda.

Code Information

UDI: (01)00884838059443(21)2086, Serial Number: 2086; UDI: (01)00884838059443(21)2099, Serial Number: 2099; UDI: (01)00884838059443(21)2095, Serial Number: 2095; UDI: (01)00884838059443(21)2097, Serial Number: 2097

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2276-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.