Z-2405-2018 Class II Terminated

Recalled by Stryker GmbH — Selzach

Recall Details

Product Type
Devices
Report Date
July 18, 2018
Initiation Date
April 27, 2018
Termination Date
May 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

Femoral Nail, RIGHT T2 GTN ¿9x320 mm, Product Number 18510932S

Reason for Recall

The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.

Distribution Pattern

The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.

Code Information

GTIN: 04546540647320; Lot code: K076A80