Z-2406-2018 Class II Terminated

Recalled by Stryker GmbH — Selzach

FDA device recall Z-2406-2018 was initiated by Stryker GmbH on Apr 27, 2018 and is designated Class II. Reason for recall: The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and… The recall status is terminated (terminated May 21, 2020).

Recall Details

Product Type
Devices
Report Date
Jul 18, 2018
Initiation Date
Apr 27, 2018
Termination Date
May 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

Femoral Nail, RIGHT T2 GTN ¿14x460 mm, Product Number 18511446S

Reason for Recall

The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.

Distribution Pattern

The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.

Code Information

GTIN: 07613252129295; Lot code: K0748D3