Z-2429-2024 Class II Ongoing

Recalled by Abbott Medical — Plano, TX

Recall Details

Product Type
Devices
Report Date
July 31, 2024
Initiation Date
May 22, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,900 units

Product Description

Abbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only

Reason for Recall

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, WA, WI, WY and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Code Information

All Serial numbers/UDI:05415067030016/05415067020246