Z-2439-2015 Class II Terminated

Recalled by PTW-Freiburg — Freiburg, N/A

FDA device recall Z-2439-2015 was initiated by PTW-Freiburg on May 27, 2015 and is designated Class II. Reason for recall: In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not… The recall status is terminated (terminated Mar 31, 2017). Affected quantity: 242 firmwares.

Recall details

Product Type
Devices
Report Date
Sep 2, 2015
Initiation Date
May 27, 2015
Termination Date
Mar 31, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
242 firmwares

Product description

PTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and health physics.

Reason for recall

In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.

Distribution pattern

Distributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.

Code information

serial numbers: **T10021: 000005, 000019, 000037, 000041, 000047, 000049, 000058, 000066, 000074, 000076, 000079, 000099, 000107, 000125, 000141, 000144, 000158, 000161, 000168, 000175, 000186, 000212, 000221, 000224, 000229, 000231, 000254, 000262, 000274, 000294, 000301, 000302, 000327, 000351, 000352, 000382, 000401, 000413, 000424, 000437, 000445, 000452, 000470, 000477, 000486, 000504, 000525, 000528, 000535, 000567, 000593, 000594, 000649, 000655, 000658, 000659, 000663, 000678, 000686, 000710, 000711, 000731, 000752, 000756, 000763, 000785, 000792, 000793, 000796, 000800, 000820, 000829, 000830, 000857, 000860, 000863, 000865 ,000871, 000874, 000890, 000892, 000946, **T10022: 000003, 000046, 000066, 000087, 000096, 000106, 000118, 000138, 000172, 000178, 000183, 000186, 000188, 000210, 000240, 000317, 000336, 000347, 000350, 000351, 000358, 000360, 000392, 000398, 000420, 000422, 000425, 000434 -000453, 000456-000469, 000474, ** T10023: 000022, 000028, 000051, 00102, 000196, 000213, 000226, 000250, 000255, 000257, 000259, 000261, 000263, 000271, 000280, 000294

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2439-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for PTW-Freiburg? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2439-2015. Report a data issue: hello@regdatalab.com.