Z-2577-2023 Class II Ongoing

Recalled by PTW-FREIBURG — Freiburg Im Breisgau

FDA device recall Z-2577-2023 was initiated by PTW-FREIBURG on Jun 7, 2023 and is designated Class II. Reason for recall: Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix… The recall status is ongoing. Affected quantity: 42 US.

Recall details

Product Type
Devices
Report Date
Sep 20, 2023
Initiation Date
Jun 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 US

Product description

BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360)

Reason for recall

Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.

Distribution pattern

US Nationwide distribution.

Code information

UDI-DI: EPTWS0800320 BeamAdjust All software Versions

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2577-2023
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.