Z-1181-2024 Class II Ongoing

Recalled by PTW-FREIBURG — Freiburg Im Breisgau, N/A

Recall Details

Product Type
Devices
Report Date
February 28, 2024
Initiation Date
February 8, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 US; 65 worldwide

Product Description

Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.

Reason for Recall

Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.

Distribution Pattern

GA

Code Information

UDI-DI: EPTWS070031 Software VERIQA (S070031), version 2.0 and 2.1