Z-2473-2012 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-2473-2012 was initiated by Zimmer, Inc. on Apr 1, 2011 and is designated Class II. Reason for recall: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging… The recall status is terminated (terminated May 16, 2014). Affected quantity: 231641 all sizes.

Recall Details

Product Type
Devices
Report Date
Oct 10, 2012
Initiation Date
Apr 1, 2011
Termination Date
May 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
231641 all sizes

Product Description

00789502202 VerSys Femoral Head Provisional 22mm +0 00789502203 VerSys Femoral Head Provisional 22mm +3.5 00789502220 VerSys Femoral Head Provisional 22mm -2 00789502230 VerSys Femoral Head Provisional 22mm +3 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Reason for Recall

Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp

Distribution Pattern

Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.

Code Information

All lots distributed from lot 69998600 (2/20/1996) to present

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2473-2012
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.