Z-2490-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
June 24, 2026
Initiation Date
May 4, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0 units

Product Description

MEDLINE Medical Procedure Kits labeled as: ORAL RESTORATION, Medline Kit Number/SKUDYNJ86253A

Reason for Recall

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Code Information

UDI/DI 10198459556654 (each), 40198459556655 (case). Lot Numbers: 25KBC063, 25JBO955.