Z-2490-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-2490-2026 was initiated by Medline Industries, LP on May 4, 2026 and is designated Class II. Reason for recall: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances… The recall status is ongoing. Affected quantity: 0 units.

Recall Details

Product Type
Devices
Report Date
Jun 24, 2026
Initiation Date
May 4, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0 units

Product Description

MEDLINE Medical Procedure Kits labeled as: ORAL RESTORATION, Medline Kit Number/SKUDYNJ86253A

Reason for Recall

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Code Information

UDI/DI 10198459556654 (each), 40198459556655 (case). Lot Numbers: 25KBC063, 25JBO955.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2490-2026
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.