Z-2542-2023 Class II

Recalled by Landauer — Glenwood, IL

FDA device recall Z-2542-2023 was initiated by Landauer on Jul 12, 2023 and is designated Class II. Reason for recall: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential… Affected quantity: 60 units.

Recall Details

Product Type
Devices
Report Date
Sep 20, 2023
Initiation Date
Jul 12, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
60 units

Product Description

nanoDot D2DNS, Model Numbers: a) 03055-1MO; b) 03055-3MO; c) 03055-OTO; d) 03055-SMO; radiation monitoring dosimeter used with the microSTAR readers

Reason for Recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Distribution Pattern

Worldwide

Code Information

UDI/DI 0860003399903, all batch numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2542-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.