Z-2568-2023 Class II

Recalled by Landauer — Glenwood, IL

FDA device recall Z-2568-2023 was initiated by Landauer on Jul 12, 2023 and is designated Class II. Reason for recall: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential… Affected quantity: 427.

Recall details

Product Type
Devices
Report Date
Sep 20, 2023
Initiation Date
Jul 12, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
427

Product description

QC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18131-000; b) 18131-KIT; c) 18131-SET; radiation monitoring dosimeter used with the microSTAR readers

Reason for recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Distribution pattern

Worldwide

Code information

UDI/DI 0860003399903, all batch numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2568-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Landauer? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2568-2023. Report a data issue: hello@regdatalab.com.