Z-2551-2023 Class II

Recalled by Landauer — Glenwood, IL

Recall Details

Product Type
Devices
Report Date
September 20, 2023
Initiation Date
July 12, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0

Product Description

QC (80 kVp) screened nanoDot D2DNS, Model Number 03066-OTO; radiation monitoring dosimeter used with the microSTAR readers

Reason for Recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Distribution Pattern

Worldwide

Code Information

UDI/DI 0860003399903, all batch numbers