FDA device recall Z-2554-2018 was initiated by RAYSEARCH LABORATORIES AB on Jun 6, 2018 and is designated Class II. Reason for recall: The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large… The recall status is terminated (terminated Apr 4, 2019). Affected quantity: 718.

Recall Details

Product Type
Devices
Report Date
Aug 1, 2018
Initiation Date
Jun 6, 2018
Termination Date
Apr 4, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
718

Product Description

RayStation Radiation Therapy Treatment Planning System, Model nos. 2.5, 3.5, 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 7.0 Product Usage RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Reason for Recall

The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must be aware to avoid incorrect dose calculations during treatment planning.

Distribution Pattern

US Nationwide distribution.

Code Information

2.5.1.89 , 3.5.0.16 , 3.5.1.6 , 4.0.0.14 , 4.0.1.4 4.0.2.9 , 4.0.3.4 , 4.3.0.14 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.9.0.42 , 5.0.0.37 , 5.0.1.11 , 5.0.2.35, 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 7.0.0.19 .

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2554-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.