Z-2563-2021 Class II Ongoing

Recalled by Elekta Inc — Saint Charles, MO

Recall Details

Product Type
Devices
Report Date
October 6, 2021
Initiation Date
September 14, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
214 units

Product Description

Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Reason for Recall

Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.

Code Information

Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10; UDI GTIN: (01)00858164002190(10) 5.40.00, (01)00858164002190(10) 5.40.01, (01)00858164002190(10) 5.40.02, (01)00858164002268(10) 5.51.10