Z-2579-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
August 8, 2018
Initiation Date
May 24, 2018
Termination Date
February 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units total

Product Description

Discovery MI Digital Ready

Reason for Recall

DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution Pattern

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Code Information

Model Number 5986030. a. Serial Number PTDMI1800003PT, System ID Number RADNETPT1839, UDI Number 01008406821209751118010421PTDMI1800003PT; b. Serial Number PTDMI1800005PT, System ID Number RADNETPT1838, UDI Number 01008406821209751118010421PTDMI1800003PT; c. Serial Number PTDMI1800009PT, System ID Number 813745DMIDR2, UDI Number 01008406821209751118021921PTDMI1800009PT