Z-2579-2018 Class II Terminated
FDA device recall Z-2579-2018 was initiated by GE Healthcare, LLC on May 24, 2018 and is designated Class II. Reason for recall: DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). The recall status is terminated (terminated Feb 8, 2019). Affected quantity: 42 units total.
Recall Details
- Product Type
- Devices
- Report Date
- Aug 8, 2018
- Initiation Date
- May 24, 2018
- Termination Date
- Feb 8, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 42 units total
Product Description
Discovery MI Digital Ready
Reason for Recall
DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Distribution Pattern
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Code Information
Model Number 5986030. a. Serial Number PTDMI1800003PT, System ID Number RADNETPT1839, UDI Number 01008406821209751118010421PTDMI1800003PT; b. Serial Number PTDMI1800005PT, System ID Number RADNETPT1838, UDI Number 01008406821209751118010421PTDMI1800003PT; c. Serial Number PTDMI1800009PT, System ID Number 813745DMIDR2, UDI Number 01008406821209751118021921PTDMI1800009PT