Z-2580-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 8, 2018
- Initiation Date
- May 24, 2018
- Termination Date
- February 8, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 42 units total
Product Description
Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring
Reason for Recall
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Distribution Pattern
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Code Information
Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI), CJRPX1800018CN (617726PET1EB), CJRPX1800021CN (404616DMI)