Z-2580-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
August 8, 2018
Initiation Date
May 24, 2018
Termination Date
February 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units total

Product Description

Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring

Reason for Recall

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution Pattern

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Code Information

Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI), CJRPX1800018CN (617726PET1EB), CJRPX1800021CN (404616DMI)