Z-2581-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-2581-2018 was initiated by GE Healthcare, LLC on May 24, 2018 and is designated Class II. Reason for recall: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). The recall status is terminated (terminated Feb 8, 2019). Affected quantity: 42 units total.

Recall Details

Product Type
Devices
Report Date
Aug 8, 2018
Initiation Date
May 24, 2018
Termination Date
Feb 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units total

Product Description

Discovery RT labeled as: a. MID BJG; b. HVY BJG

Reason for Recall

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution Pattern

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Code Information

Model Number 2374682-17. a. MID BJG. Serial Number CBCIG1800017HM (System ID 847695CTRT, UDI 01008406821186991118020021CBCIG1800017HM), Serial Number CBCIG1700070HM (System ID 251665MCIRT, UDI 01008406821186991117120021CBCIG1700070HM), Serial Number CBCIG1700064HM (System ID 973831RT, UDI 01008406821186991117120021CBCIG1700064HM), Serial Number CBCIG1800001HM (System ID 708747RT590, UDI 01008406821186991118010021CBCIG1800001HM); b. HVY BJG. Serial Number CBCIG1700062HM (System ID 904819RTCT, UDI 01008406821186991117120021CBCIG1700062HM), Serial Number CBCIG1700040HM (System ID 843692GSOCT, UDI 01008406821186991117100021CBCIG1700040HM)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2581-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.