Z-2581-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
August 8, 2018
Initiation Date
May 24, 2018
Termination Date
February 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units total

Product Description

Discovery RT labeled as: a. MID BJG; b. HVY BJG

Reason for Recall

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution Pattern

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Code Information

Model Number 2374682-17. a. MID BJG. Serial Number CBCIG1800017HM (System ID 847695CTRT, UDI 01008406821186991118020021CBCIG1800017HM), Serial Number CBCIG1700070HM (System ID 251665MCIRT, UDI 01008406821186991117120021CBCIG1700070HM), Serial Number CBCIG1700064HM (System ID 973831RT, UDI 01008406821186991117120021CBCIG1700064HM), Serial Number CBCIG1800001HM (System ID 708747RT590, UDI 01008406821186991118010021CBCIG1800001HM); b. HVY BJG. Serial Number CBCIG1700062HM (System ID 904819RTCT, UDI 01008406821186991117120021CBCIG1700062HM), Serial Number CBCIG1700040HM (System ID 843692GSOCT, UDI 01008406821186991117100021CBCIG1700040HM)