Z-2581-2018 Class II Terminated
FDA device recall Z-2581-2018 was initiated by GE Healthcare, LLC on May 24, 2018 and is designated Class II. Reason for recall: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). The recall status is terminated (terminated Feb 8, 2019). Affected quantity: 42 units total.
Recall Details
- Product Type
- Devices
- Report Date
- Aug 8, 2018
- Initiation Date
- May 24, 2018
- Termination Date
- Feb 8, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 42 units total
Product Description
Discovery RT labeled as: a. MID BJG; b. HVY BJG
Reason for Recall
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Distribution Pattern
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Code Information
Model Number 2374682-17. a. MID BJG. Serial Number CBCIG1800017HM (System ID 847695CTRT, UDI 01008406821186991118020021CBCIG1800017HM), Serial Number CBCIG1700070HM (System ID 251665MCIRT, UDI 01008406821186991117120021CBCIG1700070HM), Serial Number CBCIG1700064HM (System ID 973831RT, UDI 01008406821186991117120021CBCIG1700064HM), Serial Number CBCIG1800001HM (System ID 708747RT590, UDI 01008406821186991118010021CBCIG1800001HM); b. HVY BJG. Serial Number CBCIG1700062HM (System ID 904819RTCT, UDI 01008406821186991117120021CBCIG1700062HM), Serial Number CBCIG1700040HM (System ID 843692GSOCT, UDI 01008406821186991117100021CBCIG1700040HM)