Z-2582-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-2582-2018 was initiated by GE Healthcare, LLC on May 24, 2018 and is designated Class II. Reason for recall: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). The recall status is terminated (terminated Feb 8, 2019). Affected quantity: 42 units total.

Recall Details

Product Type
Devices
Report Date
Aug 8, 2018
Initiation Date
May 24, 2018
Termination Date
Feb 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units total

Product Description

Revolution CT 160 1.5D STD WAUK

Reason for Recall

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution Pattern

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Code Information

Model Number 5590000-6. Serial number REVVX1800021CN (System ID 812842REV, UDI 01008406821185521118020021REVVX1800021CN), Serial Number REVVX1800004CN (System ID 269375REVOCT, UDI 01008406821185521118010021REVVX1800004CN).