Z-2586-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 8, 2018
- Initiation Date
- May 24, 2018
- Termination Date
- February 8, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 42 units total
Product Description
Optima CT 540
Reason for Recall
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Distribution Pattern
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Code Information
Serial number 500925HM3, System ID Number 615320OPT