Z-2598-2018 Class III Terminated
FDA device recall Z-2598-2018 was initiated by Becton, Dickinson And Company, BD Biosciences on May 16, 2018 and is designated Class III. Reason for recall: The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use. The recall status is terminated (terminated Aug 13, 2018). Affected quantity: 8 devices.
Recall Details
- Product Type
- Devices
- Report Date
- Aug 8, 2018
- Initiation Date
- May 16, 2018
- Termination Date
- Aug 13, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8 devices
Product Description
BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA. The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.
Reason for Recall
The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
Distribution Pattern
US Distribution was made to FL, KS, KY, MA, NJ, and SD.
Code Information
Serial #2662879000001, UDI (01)00382906628794(11)170915(21)Z662879000001; Serial #2662879000002, UDI (01)00382906628794(11)170915(21)Z662879000002; Serial #2662879000003, UDI (01)00382906628794(11)170915(21)Z662879000003; Serial #2662879000004, UDI (01)00382906628794(11)170915(21)Z662879000004; Serial #2662879000009, UDI (01)00382906628794(11)170915(21)Z662879000009; Serial #2662879000010; UDI (01)00382906628794(11 )170915(21)Z662879000010; Serial #2662879000011; UDI (01)00382906628794(11)170915(21)Z662879000011; and Serial #2662879000012, UDI (01)00382906628794(11)170915(21)Z662879000012.