Z-2333-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 1, 2021
- Initiation Date
- July 22, 2021
- Termination Date
- March 24, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 23 reagents
Product Description
BD" Mouse IgG2a Isotype Control APC X39 ASR
Reason for Recall
Complaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.
Distribution Pattern
Worldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.
Code Information
Catalog Number: 340757 UDI: 00382903407576 Lot Number: 0253432