Z-0074-2020 Class II Terminated

Recalled by Becton, Dickinson And Company, BD Biosciences — San Jose, CA

FDA device recall Z-0074-2020 was initiated by Becton, Dickinson And Company, BD Biosciences on Aug 19, 2019 and is designated Class II. Reason for recall: Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio… The recall status is terminated (terminated Apr 12, 2024). Affected quantity: 22.

Recall details

Product Type
Devices
Report Date
Oct 16, 2019
Initiation Date
Aug 19, 2019
Termination Date
Apr 12, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22

Product description

BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes

Reason for recall

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

Distribution pattern

U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore

Code information

BD FACSLyric Flow Cytometer serial numbers: R663029000001, R663029000002, R663029000003, R663029000004, R663029000005, R663029000006, R663029000007, R663029000008, R663029000010, R663029000011, R663029000012, R663029000014, R663029000015, R663029000016, R663029000017, R663029000018, R663029000021, R663029000025, R663029000031, R659180000136, R659180000160, R659180000238.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0074-2020
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0074-2020. Report a data issue: hello@regdatalab.com.