Z-2640-2026 Class II Ongoing

Recalled by Accuray Incorporated — Madison, WI

FDA device recall Z-2640-2026 was initiated by Accuray Incorporated on Jun 16, 2026 and is designated Class II. Reason for recall: In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using… The recall status is ongoing. Affected quantity: 48 units.

Recall details

Product Type
Devices
Report Date
Jul 8, 2026
Initiation Date
Jun 16, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48 units

Product description

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Reason for recall

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.

Code information

All systems prior with software version 11X that utilize the Exchange Table

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2640-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2640-2026. Report a data issue: hello@regdatalab.com.