Z-1272-2021 Class II Terminated

Recalled by Accuray Incorporated — Madison, WI

FDA device recall Z-1272-2021 was initiated by Accuray Incorporated on Feb 26, 2021 and is designated Class II. Reason for recall: The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll. The recall status is terminated (terminated Jun 27, 2022). Affected quantity: 243 units.

Recall details

Product Type
Devices
Report Date
Mar 24, 2021
Initiation Date
Feb 26, 2021
Termination Date
Jun 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
243 units

Product description

CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason for recall

The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

Code information

Product Codes: 019200, 020600, 021831, 022986, 023730, 025000, 027600, 028000, 030000, 031000, 031500, 0660000, 029640-001, 029828-003, 031000-004, 032000-001, 032000-004, 033000-000, 033000-010, 053101-010, 053101-M-000, 053301-010, 053301-M-000, 053401-010, 053401-M-000, 054000-002, 054000-004, 0660000-000, 1010140-000, 1047641-000 Serial Numbers: C0063, C0067, C0008, C0076, C0077, C0085, C0098, C0101, C0112, C0126, C0131, C0135, C0138, C0151, C0155, C0162, C0166, C0172, C0174, C0176, C0177, C0186, C0187, C0188, C0195, C0197, C0200, C0206, C0208, C0211, C0213, C0228, C0231, C0234, C0240, C0242, C0246, C0253, C0263, C0266, C0271, C0239, C0441, C0446, C0447, C0456, C0460, C0464, C0465, C0469, C0472, C0475, C0481, C0506, C0507, C0193, C0241, C0293, C0090, C0092, C0093, C0100, C0103, C0104, C0115, C0117, C0119, C0124, C0130, C0148, C0150, C0153, C0168, C0181, C0182, C0194, C0205, C0226, C0235, C0236, C0244, C0254, C0255, C0258, C0259, C0267, C0268, C0274, C0276, C0280, C0282, C0283, C0285, C0286, C0289, C0291, C0292, C0297, C0298, C0299, C0300, C0301, C0302, C0303, C0304, C0305, C0306, C0309, C0310, C0312, C0313, C0317, C0319, C0321, C0325, C0328, C0329, C0331, C0339, C0347, C0353, C0357, C0365, C0377, C0382, C0386, C0388, C0399, C0401, C0403, C0405, C0416, C0432, C0435, C0451, C0064, C0071, C0074, C0091, C0099, C0102, C0137, C0238, C0249, C0252, C0270, C0277, C0294, C0417, C0476, C0384, C0414, C0424, C0439, C0427, C0428, C0344, C0423, C0372, C0387, C0330, C0343, C0345, C0346, C0348, C0351, C0352, C0356, C0360, C0363, C0368, C0369, C0370, C0373, C0380, C0381, C0383, C0389, C0390, C0394, C0395, C0396, C0397, C0398, C0402, C0404, C0406, C0411, C0413, C0415, C0420, C0422, C0429, C0430, C0436, C0437, C0438, C0440, C0444, C0445, C0448, C0453, C0457, C0458, C0466, C0461, C0467, C0473, C0480, C0482, C0484, C0486, C0487, C0488, C0489, C0490, C0491, C0492, C0493, C0494, C0495, C0496, C0497, C0500, C0505, C0513, C0014, C0022, C0040, C0042, C0046, C0047, C0056, C0059, C0066, C0079, C0086, C0089, C0140, C0173, C0311, C0355, C0470

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1272-2021
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Accuray Incorporated? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1272-2021. Report a data issue: hello@regdatalab.com.