Z-2489-2023 Class II Terminated

Recalled by Accuray Incorporated — Madison, WI

FDA device recall Z-2489-2023 was initiated by Accuray Incorporated on Aug 3, 2023 and is designated Class II. Reason for recall: It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the… The recall status is terminated (terminated Oct 30, 2026). Affected quantity: 82.

Recall details

Product Type
Devices
Report Date
Sep 6, 2023
Initiation Date
Aug 3, 2023
Termination Date
Oct 30, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
82

Product description

Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.

Reason for recall

It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.

Distribution pattern

Worldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ, OR, PA, SC, SD, TX, and WA. There was government distribution and no military distribution. The countries of Australia, Belgium, Canada, Colombia, France, India, Indonesia, Italy, Japan, Kazakhstan, Malaysia, Poland, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.

Code information

UDI-DI number 00811376030009. Serial numbers 4010505, 4010506, 4010507, 4010508, 4010509, 4010510, 4010511, 4010513, 4010515, 4010516, 4010517, 4010518, 4010519, 4010521, 4010522, 4010523, 4010524, 4010525, 4010526, 4010527, 4010528, 4010529, 4010530, 4010531, 4010532, 4010533, 4010536, 4010537, 4010538, 4010539, 4010541, 4010542, 4010543, 4010544, 4010545, 4010546, 4010548, 4010549, 4010550, 4010551, 4010552, 4010553, 4010554, 4010555, 4010556, 4010557, 4010558, 4010559, 4010560, 4010561, 4010562, 4010563, 4010564, 4010565, 4010566, 4010567, 4010568, 4010569, 4010570, 4010571, 4010573, 4010574, 4010577, 4010579, 4010582, 4010583, 4010584, 4010585, 4010586, 4010587, 4010588, 4010589, 4010590, 4010592, 4010593, 4010594, 4010596, 4010599, 4010607, 4010600, 4010601, and 4010609.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2489-2023
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Accuray Incorporated? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2489-2023. Report a data issue: hello@regdatalab.com.