Z-2649-2016 Class II Terminated

Recalled by Bovie Medical Corporation — Clearwater, FL

FDA device recall Z-2649-2016 was initiated by Bovie Medical Corporation on Jul 18, 2016 and is designated Class II. Reason for recall: The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip. The recall status is terminated (terminated Dec 14, 2016). Affected quantity: 95.

Recall Details

Product Type
Devices
Report Date
Aug 31, 2016
Initiation Date
Jul 18, 2016
Termination Date
Dec 14, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
95

Product Description

BVX-450NR Bovie Precise 360 Handpiece 45cm, needle Product is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.

Reason for Recall

The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of: MD, GA, MO and FL and the country of Finland.

Code Information

Lot 0316D.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2649-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.