Z-2735-2024 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

Recall Details

Product Type
Devices
Report Date
September 4, 2024
Initiation Date
May 22, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
183 analyzers

Product Description

DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

Reason for Recall

When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom of Great Britain and Northern Ireland.

Code Information

Software versions 1.17.0 or below, UDI-DI 15099590732103.