Recalled by Degania Silicone, Ltd. — Dganya Bet, N/A

FDA device recall Z-2761-2018 was initiated by Degania Silicone, Ltd. on Jun 21, 2018 and is designated Class II. Reason for recall: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient. The recall status is terminated (terminated Oct 25, 2024). Affected quantity: 16133.

Recall Details

Product Type
Devices
Report Date
Sep 26, 2018
Initiation Date
Jun 21, 2018
Termination Date
Oct 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16133

Product Description

Temperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Reason for Recall

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Distribution Pattern

Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.

Code Information

V17018057, V17022043, V17023777, V17018058, V17012115, V17002305, V17002306, V17002308

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2761-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.