Z-2763-2019 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-2763-2019 was initiated by Beckman Coulter Inc. on Sep 18, 2018 and is designated Class II. Reason for recall: Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%). The recall status is terminated (terminated Feb 21, 2024). Affected quantity: Access Cortisol Reagent Assay.

Recall Details

Product Type
Devices
Report Date
Oct 9, 2019
Initiation Date
Sep 18, 2018
Termination Date
Feb 21, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Access Cortisol Reagent Assay

Product Description

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Reason for Recall

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.

Code Information

Lot Number 831755.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2763-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.