Z-2774-2018 Class II Terminated

Recalled by Degania Silicone, Ltd. — Dganya Bet, N/A

FDA device recall Z-2774-2018 was initiated by Degania Silicone, Ltd. on Jun 21, 2018 and is designated Class II. Reason for recall: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient. The recall status is terminated (terminated Oct 25, 2024). Affected quantity: 500.

Recall Details

Product Type
Devices
Report Date
Sep 26, 2018
Initiation Date
Jun 21, 2018
Termination Date
Oct 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
500

Product Description

Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101880MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Reason for Recall

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Distribution Pattern

Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.

Code Information

S17003321

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2774-2018
openFDA file date
Oct 23, 2026