Z-2779-2015 Class III Terminated
FDA device recall Z-2779-2015 was initiated by Leica Microsystems, Inc. on Aug 5, 2015 and is designated Class III. Reason for recall: The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure. The recall status is terminated (terminated Jan 5, 2017). Affected quantity: 901.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 23, 2015
- Initiation Date
- Aug 5, 2015
- Termination Date
- Jan 5, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 901
Product Description
Reason for Recall
Distribution Pattern
Worldwide Distribution - US including U.S. including AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NJ, NM, NY, OH, OR, PA, TX, VA, WA, WI, WV and District of Columbia. Internationall to AE, AT, AU , BD , BE, BF, BG,BR, CA, CL, CN, CO, CR, DE, DK, EC, EG, ES, ET, FI, FR, GB, HU, ID, IN, IT, JM, JP, KR, LK, MA, MT, MY, NL, NZ, PA, PE, PH, PL, PT, RO, RU, SA, SE, SG, TH, TR, TW, VN and ZA.