Z-2980-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 11, 2024
- Initiation Date
- July 17, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
Reason for Recall
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.
Code Information
UDI: (01)15099590673567/ All serial numbers