Z-2979-2024 Class II Ongoing

Recalled by Beckman Coulter Biomedical GmbH — Munchen

FDA device recall Z-2979-2024 was initiated by Beckman Coulter Biomedical GmbH on Jul 17, 2024 and is designated Class II. Reason for recall: The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Sep 11, 2024
Initiation Date
Jul 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated

Product description

PE Centrifuge, REF: B36365, a component of the Power Express

Reason for recall

The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.

Code information

UDI: (01)15099590673550/ All serial numbers Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2979-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Beckman Coulter Biomedical GmbH? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Beckman Coulter Biomedical GmbH: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Beckman Coulter Biomedical GmbH →

Source: openFDA record for Z-2979-2024. Report a data issue: hello@regdatalab.com.