Z-3180-2024 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

FDA device recall Z-3180-2024 was initiated by Beckman Coulter, Inc. on Jul 10, 2024 and is designated Class II. Reason for recall: Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing… The recall status is ongoing. Affected quantity: 170 units.

Recall Details

Product Type
Devices
Report Date
Sep 25, 2024
Initiation Date
Jul 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
170 units

Product Description

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument

Reason for Recall

Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

Code Information

UDI/DI 15099590230623, Serial Numbers: 300116, 300117, 300118, 300123, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, 300156, 300157, 300158, 300160, 300161, 300162, 300163, 300164, 300166, 300167, 300168, 300169, 300170, 300171, 300172, 300174, 300176, 300178, 300179, 300180, 300181, 300182, 300183, 300188, 300189, 300190, 300192, 300193, 300194, 300195, 300196, 300197, 300198, 300199, 300200, 300201, 300202, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212, 300213, 300214, 300215, 300216, 300217, 300218, 300219, 300220, 300221, 300222, 300223, 300224, 300226, 300227, 300228, 300229, 300230, 300231, 300232, 300233, 300234, 300235, 300236, 300237, 300238, 300239, 300240, 300241, 300242, 300243, 300244, 300245, 300246, 300247, 300248, 300249, 300251, 300252, 300253, 300254, 300255, 300256, 300257, 300258, 300260, 300261, 300263, 300264, 300266, 300269, 300270, 300271, 300272, 300273, 300274, 300275, 300276, 300277, 300280, 300281, 300282, 300284, 300285, 300286, 300287, 300291, 300292, 300293, 300294, 300295, 300297, 300307, 300308, 300310, 300312, 300313, 300314, 300315, 300316, 300317, 300319, 300320, 300321, 300322, 300323, 300324, 300325, 300326, 300327, 300328, 300329, 300330, 300332, 300333

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3180-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.