Z-3184-2018 Class III Terminated

Recalled by Ortho-Clinical Diagnostics — Rochester, NY

Recall Details

Product Type
Devices
Report Date
September 26, 2018
Initiation Date
July 11, 2018
Termination Date
August 11, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
168919

Product Description

VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage: VITROS Cl Slides (Product Code 684 4471) - For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Reason for Recall

The slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for urine samples. The measuring range (15-300 mmol/L) for urine samples when tested with the Cl- slide is not affected. The current Cl- LoQ is 5 mmol/L and the LoQ determined by this investigation is 15 mmol/L.

Distribution Pattern

Worldwide Distribution - US Nationwide- The products were distributed to the following foreign countries: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.

Code Information

GEN 10 and above